Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 73: (1 year) CAL VER /5520 DATE: 16-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 18 NUMBER OF LINES: 100 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 70 Series service manual (PN 731034). # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # 73_dsp.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Allow the UUT to stablize to room temperature 1.005 DISP 23degC +/-5degC (73degF +/-9degF). 1.005 DISP 1.005 DISP Check the fuses and battery, and replace them if 1.005 DISP necessary. 1.006 HEAD {DISPLAY TEST} 1.007 PICE 73_DSP 2.001 HEAD {RESISTANCE TESTS} 2.002 DISP Select the Ohms Function. 2.003 DISP Connect the 5520 to the UUT as follows: 2.003 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 2.003 DISP [32] 5520A NORMAL LO to UUT COM 2.003 DISP [32] 5520A AUX HI to UUT V/Ohm/Diode 2.003 DISP [32] 5520A AUX LO to UUT COM 2.004 HEAD RESISTANCE TESTS: {320 Ohm Range} 2.005 5520 320 100.0Z 1.2U CW 3.001 HEAD RESISTANCE TESTS: {3200 Ohm Range} 3.002 5520 3200 1000Z 11U CW 4.001 HEAD RESISTANCE TESTS: {32 kOhm Range} 4.002 5520 32 10.00kZ 0.11U CW 5.001 HEAD RESISTANCE TESTS: {320 kOhm Range} 5.002 DISP Connect the 5520 to the UUT as follows: 5.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 5.002 DISP [32] 5520A NORMAL LO to UUT COM 5.003 5520 320 100.0kZ 1.1U 2W 6.001 HEAD RESISTANCE TESTS: {3.2 MOhm Range} 6.002 5520 3.2 1.000MZ 0.011U 2W 7.001 HEAD RESISTANCE TESTS: {32 MOhm Range} 7.002 5520 32 10.00MZ 0.31U 2W 8.001 HEAD {DC MILLIVOLT TEST} 8.002 DISP Select the 300mV DC Function. 8.003 HEAD DC MILLIVOLT TEST: {320mV Range} 8.004 5520 320 300.0mV 2.2U 2W 9.001 HEAD {DC VOLTAGE TEST} 9.002 DISP Select the V DC Function. 9.003 HEAD DC VOLTAGE TEST: {3.2V Range} 9.004 5520 3.2 2.700V 0.02U 2W 10.001 HEAD DC VOLTAGE TEST: {32V Range} 10.002 5520 32 27.00V 0.2U 2W 11.001 HEAD DC VOLTAGE TEST: {320V Range} 11.002 5520 320 270.0V 2.0U 2W 12.001 HEAD DC VOLTAGE TEST: {1000V Range} 12.002 5520 1000 1000V 9U 2W 13.001 HEAD {AC VOLTAGE TESTS} 13.002 DISP Select the V AC Function. 13.003 HEAD AC VOLTAGE TEST: {3.2V Range} 13.004 5520 3.2 2.700V 0.083U 100H SI 2W 14.001 5520 3.2 2.700V 0.083U 500H SI 2W 15.001 HEAD AC VOLTAGE TEST: {750V Range} 15.002 5520 750 750V 25U 100H SI 2W 16.001 5520 750 750V 25U 1000H SI 2W 17.001 HEAD {DC CURRENT TESTS} 17.002 DISP Select the A DC Function. 17.003 DISP Connect the 5520 to the UUT as follows: 17.003 DISP [32] 5520A AUX HI to UUT 10A 17.003 DISP [32] 5520A AUX LO to UUT COM 17.004 HEAD DC CURRENT TESTS: {10A Range} 17.005 5520 10 2.00A 0.07U 2W 18.001 END